CVE-2014-5432
Last modified
CVE-2014-5432 is a vulnerability of currently unknown severity. Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. EPSS estimates a 2.57% chance of exploitation in the next 30 days.
Description
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. Baxter asserts that this vulnerability only allows access to features and functionality on the WBM and that the SIGMA Spectrum infusion pump cannot be controlled from the WBM. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
Metrics
Weakness Enumeration
Affected Software
| Vendor | Product | Versions |
|---|---|---|
| Baxter | Sigma Spectrum Infusion System Firmware | 6.05 |
References
- https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01Mitigation, Third Party Advisory, US Government Resource
- https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01Mitigation, Third Party Advisory, US Government Resource
Timeline
- Published
- Last Modified
- Status
- Modified
Frequently Asked Questions
What is CVE-2014-5432?
How severe is CVE-2014-5432?
How do I fix CVE-2014-5432?
Are you affected by CVE-2014-5432?
Run a free Strix scan to check your systems for this vulnerability.
Scan your code nowSource: NVD / NIST