CVE-2017-5149
Last modified
CVE-2017-5149 is a high-severity vulnerability rated 8.9/10 on the CVSS scale. An issue was discovered in St. Jude Medical Merlin@home, versions prior to Version 8.2.2 (RF models: EX1150; Inductive models: EX1100; and Inductive models: EX1100 with MerlinOnDemand capability). EPSS estimates a 0.96% chance of exploitation in the next 30 days.
Description
An issue was discovered in St. Jude Medical Merlin@home, versions prior to Version 8.2.2 (RF models: EX1150; Inductive models: EX1100; and Inductive models: EX1100 with MerlinOnDemand capability). The identities of the endpoints for the communication channel between the transmitter and St. Jude Medical's web site, Merlin.net, are not verified. This may allow a man-in-the-middle attacker to access or influence communications between the identified endpoints.
Metrics
CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:L/I:H/A:H
Weakness Enumeration
Affected Software
| Vendor | Product | Versions |
|---|---|---|
| Abbott | Merlin\@Home Firmware | <= 8.0 |
References
- http://www.securityfocus.com/bid/95331Third Party Advisory, VDB Entry
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-009-01AMitigation, Third Party Advisory, US Government Resource
- http://www.securityfocus.com/bid/95331Third Party Advisory, VDB Entry
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-009-01AMitigation, Third Party Advisory, US Government Resource