CVE-2018-10597
Last modified
CVE-2018-10597 is a high-severity vulnerability rated 8.3/10 on the CVSS scale. IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that allows an unauthenticated attacker to access memory ("write-what-where") from an attacker-chosen device address within the same subnet.. EPSS estimates a 0.39% chance of exploitation in the next 30 days.
Description
IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that allows an unauthenticated attacker to access memory ("write-what-where") from an attacker-chosen device address within the same subnet.
Metrics
CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:H/A:H
Weakness Enumeration
Affected Software
| Vendor | Product | Versions |
|---|---|---|
| Philips | Intellivue Mp2 Firmware | All versions |
| Philips | Intellivue X2 Firmware | All versions |
| Philips | Intellivue Mp30 Firmware | All versions |
| Philips | Intellivue Mp50 Firmware | All versions |
| Philips | Intellivue Mp70 Firmware | All versions |
| Philips | Intellivue Np90 Firmware | All versions |
| Philips | Intellivue Mx700 Firmware | All versions |
| Philips | Intellivue Mx800 Firmware | All versions |
| Philips | Intellivue Mx400 Firmware | All versions |
| Philips | Intellivue Mx450 Firmware | All versions |
| Philips | Intellivue Mx500 Firmware | All versions |
| Philips | Intellivue Mx550 Firmware | All versions |
| Philips | Intellivue X3 Firmware | All versions |
| Philips | Intellivue Mx100 Firmware | All versions |
| Philips | Avalon Fetal\/Maternal Monitors Fm20 Firmware | All versions |
| Philips | Avalon Fetal\/Maternal Monitors Fm30 Firmware | All versions |
| Philips | Avalon Fetal\/Maternal Monitors Fm40 Firmware | All versions |
| Philips | Avalon Fetal\/Maternal Monitors Fm50 Firmware | All versions |
References
- https://ics-cert.us-cert.gov/advisories/ICSMA-18-156-01Third Party Advisory, US Government Resource
- https://ics-cert.us-cert.gov/advisories/ICSMA-18-156-01Third Party Advisory, US Government Resource
Timeline
- Published
- Last Modified
- Status
- Modified
Frequently Asked Questions
What is CVE-2018-10597?
How severe is CVE-2018-10597?
How do I fix CVE-2018-10597?
Are you affected by CVE-2018-10597?
Run a free Strix scan to check your systems for this vulnerability.
Scan your code nowSource: NVD / NIST
