CVE-2020-25183

HIGHCVSS 8.8/10EPSS 0.79%

Last modified

CVE-2020-25183 is a high-severity vulnerability rated 8.8/10 on the CVSS scale. Medtronic MyCareLink Smart 25000 contains an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an attacker to use another mobile device or malicious application on the patient’s smartphone to authenticate to the patient’s Medtronic Smart Reader, fooling the device into believing it is communicating with the original Medtronic smart phone application when executed within range of Bluetooth communication.. EPSS estimates a 0.79% chance of exploitation in the next 30 days.

Description

Medtronic MyCareLink Smart 25000 contains an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an attacker to use another mobile device or malicious application on the patient’s smartphone to authenticate to the patient’s Medtronic Smart Reader, fooling the device into believing it is communicating with the original Medtronic smart phone application when executed within range of Bluetooth communication.

Metrics

CVSS 3.1
8.8/10

CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

EPSS Probability
0.79%

51.5th percentile

Probability of exploitation in the next 30 days. Learn more

Weakness Enumeration

Affected Software

VendorProductVersions
MedtronicMycarelink Smart Model 25000 FirmwareAll versions

References

Timeline

Published
Last Modified
Status
Modified

Frequently Asked Questions

What is CVE-2020-25183?
Medtronic MyCareLink Smart 25000 contains an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an attacker to use another mobile device or malicious application on the patient’s smartphone to authenticate to the patient’s Medtronic Smart Reader, fooling the device into believing it is communicating with the original Medtronic smart phone application when executed within range of Bluetooth communication.
How severe is CVE-2020-25183?
CVE-2020-25183 has a CVSS score of 8.8/10 (HIGH severity). The EPSS model estimates a 0.79% probability of exploitation in the next 30 days.
How do I fix CVE-2020-25183?
Check the vendor references and advisories linked above for patched versions and mitigation guidance. You can also run a Strix scan to test if your systems are affected.

Are you affected by CVE-2020-25183?

Run a free Strix scan to check your systems for this vulnerability.

Scan your code now

Source: NVD / NIST