CVE-2020-27282
Last modified
CVE-2020-27282 is a medium-severity vulnerability rated 4.3/10 on the CVSS scale. In Hamilton Medical AG,T1-Ventillator versions 2.2.3 and prior, an XML validation vulnerability in the ventilator allows privileged attackers with physical access to render the device persistently unusable by uploading specially crafted configuration files.. EPSS estimates a 0.25% chance of exploitation in the next 30 days.
Description
In Hamilton Medical AG,T1-Ventillator versions 2.2.3 and prior, an XML validation vulnerability in the ventilator allows privileged attackers with physical access to render the device persistently unusable by uploading specially crafted configuration files.
Metrics
CVSS:3.1/AV:P/AC:L/PR:L/UI:N/S:U/C:N/I:N/A:H
Weakness Enumeration
Affected Software
| Vendor | Product | Versions |
|---|---|---|
| Hamilton-Medical | Hamilton-T1 Firmware | <= 2.2.3 |
References
- https://us-cert.cisa.gov/ics/advisories/icsma-21-047-01Third Party Advisory, US Government Resource
- https://us-cert.cisa.gov/ics/advisories/icsma-21-047-01Third Party Advisory, US Government Resource
Timeline
- Published
- Last Modified
- Status
- Modified
Frequently Asked Questions
What is CVE-2020-27282?
How severe is CVE-2020-27282?
How do I fix CVE-2020-27282?
Are you affected by CVE-2020-27282?
Run a free Strix scan to check your systems for this vulnerability.
Scan your code nowSource: NVD / NIST