CVE-2023-0888
Last modified
CVE-2023-0888 is a high-severity vulnerability rated 7.2/10 on the CVSS scale. An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. EPSS estimates a 0.60% chance of exploitation in the next 30 days.
Description
An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks
Metrics
CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H
Weakness Enumeration
Affected Software
| Vendor | Product | Versions |
|---|---|---|
| Bbraun | Battery-Pack Sp With Wifi Firmware | <= 053l000092 |
| Bbraun | Battery-Pack Sp With Wifi Firmware | <= 054u000092 |
References
Timeline
- Published
- Last Modified
- Status
- Modified
Frequently Asked Questions
What is CVE-2023-0888?
How severe is CVE-2023-0888?
How do I fix CVE-2023-0888?
Are you affected by CVE-2023-0888?
Run a free Strix scan to check your systems for this vulnerability.
Scan your code nowSource: NVD / NIST