CVE-2023-0888 is a high-severity vulnerability rated 7.2/10 on the CVSS scale. An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. EPSS estimates a 0.60% chance of exploitation in the next 30 days.

Description

An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks

Metrics

CVSS 3.1
7.2/10

CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H

EPSS Probability
0.60%

44.3th percentile

Probability of exploitation in the next 30 days. Learn more

Weakness Enumeration

Affected Software

VendorProductVersions
BbraunBattery-Pack Sp With Wifi Firmware<= 053l000092
BbraunBattery-Pack Sp With Wifi Firmware<= 054u000092

References

Timeline

Published
Last Modified
Status
Modified

Frequently Asked Questions

What is CVE-2023-0888?
An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks
How severe is CVE-2023-0888?
CVE-2023-0888 has a CVSS score of 7.2/10 (HIGH severity). The EPSS model estimates a 0.60% probability of exploitation in the next 30 days.
How do I fix CVE-2023-0888?
Check the vendor references and advisories linked above for patched versions and mitigation guidance. You can also run a Strix scan to test if your systems are affected.

Are you affected by CVE-2023-0888?

Run a free Strix scan to check your systems for this vulnerability.