CVE-2023-25931
Last modified
CVE-2023-25931 is a medium-severity vulnerability rated 6.8/10 on the CVSS scale. Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. EPSS estimates a 0.27% chance of exploitation in the next 30 days.
Description
Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.
Metrics
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Weakness Enumeration
Affected Software
| Vendor | Product | Versions |
|---|---|---|
| Medtronic | Interstim X Clinician | a51300 |
| Medtronic | Micro Clinician | a51200 |
References
Timeline
- Published
- Last Modified
- Status
- Modified
Frequently Asked Questions
What is CVE-2023-25931?
How severe is CVE-2023-25931?
How do I fix CVE-2023-25931?
Are you affected by CVE-2023-25931?
Run a free Strix scan to check your systems for this vulnerability.
Scan your code nowSource: NVD / NIST