CVE-2025-14096

HIGHCVSS 8.4/10EPSS 0.12%

Last modified

CVE-2025-14096 is a high-severity vulnerability rated 8.4/10 on the CVSS scale. A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system. Other related CVE's are CVE-2025-14095 & CVE-2025-14097. Affected customers have been informed about this vulnerability. EPSS estimates a 0.12% chance of exploitation in the next 30 days.

Description

A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system. Other related CVE's are CVE-2025-14095 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Attacker requires physical access to the analyzer. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.

Metrics

CVSS 3.1
8.4/10

CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

EPSS Probability
0.12%

1.9th percentile

Probability of exploitation in the next 30 days. Learn more

Weakness Enumeration

References

Timeline

Published
Last Modified
Status
Deferred

Frequently Asked Questions

What is CVE-2025-14096?
A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system. Other related CVE's are CVE-2025-14095 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Attacker requires physical access to the analyzer. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.
How severe is CVE-2025-14096?
CVE-2025-14096 has a CVSS score of 8.4/10 (HIGH severity). The EPSS model estimates a 0.12% probability of exploitation in the next 30 days.
How do I fix CVE-2025-14096?
Check the vendor references and advisories linked above for patched versions and mitigation guidance. You can also run a Strix scan to test if your systems are affected.

Are you affected by CVE-2025-14096?

Run a free Strix scan to check your systems for this vulnerability.

Scan your code now

Source: NVD / NIST