CVE-2026-7251

CRITICALCVSS 9.8/10EPSS 0.50%

Last modified

CVE-2026-7251 is a critical-severity vulnerability rated 9.8/10 on the CVSS scale. Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. EPSS estimates a 0.50% chance of exploitation in the next 30 days.

Description

Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.

Metrics

CVSS 3.1
9.8/10

CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

CVSS 4.0
9.3/10

CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X

EPSS Probability
0.50%

39.7th percentile

Probability of exploitation in the next 30 days. Learn more

Weakness Enumeration

Affected Software

Source: CNA advisory (CVE.org). NVD analysis pending.

VendorProductVersions
EppendorfBioFlo 320All

References

Timeline

Published
Last Modified
Status
Awaiting Analysis

Frequently Asked Questions

What is CVE-2026-7251?
Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.
How severe is CVE-2026-7251?
CVE-2026-7251 has a CVSS score of 9.8/10 (CRITICAL severity). The EPSS model estimates a 0.50% probability of exploitation in the next 30 days.
How do I fix CVE-2026-7251?
Check the vendor references and advisories linked above for patched versions and mitigation guidance. You can also run a Strix scan to test if your systems are affected.

How Strix Helps

Related CVEs from 2026

Are you affected by CVE-2026-7251?

Run a free Strix scan to check your systems for this vulnerability.

Scan your code now

Source: NVD / NIST