CVE-2019-6538
Last modified
CVE-2019-6538 is a medium-severity vulnerability rated 6.5/10 on the CVSS scale. The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. EPSS estimates a 0.84% chance of exploitation in the next 30 days.
Description
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.
Metrics
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N
Weakness Enumeration
Affected Software
| Vendor | Product | Versions |
|---|---|---|
| Medtronic | Mycarelink Monitor Firmware | 24950 |
| Medtronic | Mycarelink Monitor Firmware | 24952 |
| Medtronic | Carelink Monitor Firmware | 2490c |
| Medtronic | Carelink 2090 Firmware | All versions |
| Medtronic | Amplia Crt-D Firmware | All versions |
| Medtronic | Claria Crt-D Firmware | All versions |
| Medtronic | Compia Crt-D Firmware | All versions |
| Medtronic | Concerto Crt-D Firmware | All versions |
| Medtronic | Concerto Ii Crt-D Firmware | All versions |
| Medtronic | Consulta Crt-D Firmware | All versions |
| Medtronic | Evera Icd Firmware | All versions |
| Medtronic | Maximo Ii Crt-D And Lcd Firmware | All versions |
| Medtronic | Mirro Icd Firmware | All versions |
| Medtronic | Nayamed Nd Icd Firmware | All versions |
| Medtronic | Primo Icd Firmware | All versions |
| Medtronic | Protecta Icd And Crt-D Firmware | All versions |
| Medtronic | Secura Icd Firmware | All versions |
| Medtronic | Virtuoso Icd Firmware | All versions |
| Medtronic | Virtuoso Ii Icd Firmware | All versions |
| Medtronic | Visia Af Icd Firmware | All versions |
| Medtronic | Viva Crt-D Firmware | All versions |
References
- http://www.securityfocus.com/bid/107544Third Party Advisory, VDB Entry
- https://ics-cert.us-cert.gov/advisories/ICSMA-19-080-01Mitigation, Third Party Advisory, US Government Resource
- http://www.securityfocus.com/bid/107544Third Party Advisory, VDB Entry
- https://ics-cert.us-cert.gov/advisories/ICSMA-19-080-01Mitigation, Third Party Advisory, US Government Resource
Timeline
- Published
- Last Modified
- Status
- Modified
Frequently Asked Questions
What is CVE-2019-6538?
How severe is CVE-2019-6538?
How do I fix CVE-2019-6538?
Are you affected by CVE-2019-6538?
Run a free Strix scan to check your systems for this vulnerability.
Scan your code nowSource: NVD / NIST