CVE-2020-15486
Last modified
CVE-2020-15486 is a medium-severity vulnerability rated 6.5/10 on the CVSS scale. An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. EPSS estimates a 0.51% chance of exploitation in the next 30 days.
Description
An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity.
Metrics
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
Affected Software
| Vendor | Product | Versions |
|---|---|---|
| Drtrust | Electrocardiogram Pen Firmware | 2.00.08 |
References
Timeline
- Published
- Last Modified
- Status
- Modified
Frequently Asked Questions
What is CVE-2020-15486?
How severe is CVE-2020-15486?
How do I fix CVE-2020-15486?
Are you affected by CVE-2020-15486?
Run a free Strix scan to check your systems for this vulnerability.
Scan your code nowSource: NVD / NIST