CVE-2020-7590

MEDIUMCVSS 6.8/10EPSS 0.35%

Last modified

CVE-2020-7590 is a medium-severity vulnerability rated 6.8/10 on the CVSS scale. A vulnerability has been identified in DCA Vantage Analyzer (All versions < V4.5 are affected by CVE-2020-7590. In addition, serial numbers < 40000 running software V4.4.0 are also affected by CVE-2020-15797). EPSS estimates a 0.35% chance of exploitation in the next 30 days.

Description

A vulnerability has been identified in DCA Vantage Analyzer (All versions < V4.5 are affected by CVE-2020-7590. In addition, serial numbers < 40000 running software V4.4.0 are also affected by CVE-2020-15797). Affected devices use a hard-coded password to protect the onboard database. This could allow an attacker to read and or modify the onboard database. Successful exploitation requires direct physical access to the device.

Metrics

CVSS 3.1
6.8/10

CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

EPSS Probability
0.35%

26.7th percentile

Probability of exploitation in the next 30 days. Learn more

Weakness Enumeration

Affected Software

VendorProductVersions
SiemensDca Vantage Analyzer Firmware< 4.5.0.0

References

Timeline

Published
Last Modified
Status
Modified

Frequently Asked Questions

What is CVE-2020-7590?
A vulnerability has been identified in DCA Vantage Analyzer (All versions < V4.5 are affected by CVE-2020-7590. In addition, serial numbers < 40000 running software V4.4.0 are also affected by CVE-2020-15797). Affected devices use a hard-coded password to protect the onboard database. This could allow an attacker to read and or modify the onboard database. Successful exploitation requires direct physical access to the device.
How severe is CVE-2020-7590?
CVE-2020-7590 has a CVSS score of 6.8/10 (MEDIUM severity). The EPSS model estimates a 0.35% probability of exploitation in the next 30 days.
How do I fix CVE-2020-7590?
Check the vendor references and advisories linked above for patched versions and mitigation guidance. You can also run a Strix scan to test if your systems are affected.

Are you affected by CVE-2020-7590?

Run a free Strix scan to check your systems for this vulnerability.

Scan your code now