CVE-2017-12718 is a vulnerability of currently unknown severity. A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. EPSS estimates a 13.02% chance of exploitation in the next 30 days.

Description

A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. The pump receives the potentially malicious input infrequently and under certain conditions, increasing the difficulty of exploitation.

Metrics

EPSS Probability
13.02%

95.8th percentile

Probability of exploitation in the next 30 days. Learn more

Weakness Enumeration

Affected Software

VendorProductVersions
Smiths-MedicalMedfusion 4000 Wireless Syringe Infusion Pump1.1
Smiths-MedicalMedfusion 4000 Wireless Syringe Infusion Pump1.5
Smiths-MedicalMedfusion 4000 Wireless Syringe Infusion Pump1.6

References

Timeline

Published
Last Modified
Status
Modified

Frequently Asked Questions

What is CVE-2017-12718?
A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. The pump receives the potentially malicious input infrequently and under certain conditions, increasing the difficulty of exploitation.
How severe is CVE-2017-12718?
Severity scoring for CVE-2017-12718 is pending analysis. The EPSS model estimates a 13.02% probability of exploitation in the next 30 days.
How do I fix CVE-2017-12718?
Check the vendor references and advisories linked above for patched versions and mitigation guidance. You can also run a Strix scan to test if your systems are affected.