CVE-2019-25719
Last modified
CVE-2019-25719 is a high-severity vulnerability rated 8.8/10 on the CVSS scale. Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.. EPSS estimates a 0.13% chance of exploitation in the next 30 days.
Description
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.
Metrics
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H
CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
Weakness Enumeration
Affected Software
Source: CNA advisory (CVE.org). NVD analysis pending.
| Vendor | Product | Versions |
|---|---|---|
| Dräger | Infinity Acute Care System | <= VG4.1.1; <= VG4.0.3 |
| Dräger | Standalone Infinity M540 patient monitor | <= VG4.1.1 |
References
Timeline
- Published
- Last Modified
- Status
- Awaiting Analysis
Frequently Asked Questions
What is CVE-2019-25719?
How severe is CVE-2019-25719?
How do I fix CVE-2019-25719?
How Strix Helps
- How Strix found a critical auth bypass in etcdStrix autonomously discovered a critical authentication bypass in etcd, later designated CVE-2026-33413.
- Autonomous PentestingAI agents that find and validate exploitable vulnerabilities like this one across your applications.
- PR ReviewsPentest every pull request so vulnerable code is caught before it ships to production.
- AI Penetration TestingHow AI-driven penetration testing continuously covers your attack surface.
Related CVEs from 2019
- CVE-2019-25712BlueAuditor 1.7.2.0 contains a buffer overflow vulnerability…5.5
- CVE-2019-25713MyT-PM 1.5.1 contains an SQL injection vulnerability that al…8.1
- CVE-2019-25714Seeyon OA A8 contains an unauthenticated arbitrary file writ…9.3
- CVE-2019-25716Dräger Infinity Delta, Delta XL, and Kappa patient monitors …7.5
- CVE-2019-25717Dräger Infinity Delta, Delta XL, and Kappa patient monitors …5.3
- CVE-2019-25718Dräger Infinity Explorer C700 contains a privilege escalatio…7.8
- CVE-2019-2572Vulnerability in the Oracle SOA Suite component of Oracle Fu…
- CVE-2019-25720Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000,…7.1
- CVE-2019-25721Dräger Infinity M300 patient worn monitors with software ver…7.1
- CVE-2019-25722Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000,…7.6
- CVE-2019-25723Dräger Perseus A500 software versions 2.00 through 2.02 cont…6.3
- CVE-2019-25724Dräger Infinity M300 patient worn monitors with software ver…7.1
Are you affected by CVE-2019-25719?
Run a free Strix scan to check your systems for this vulnerability.
Scan your code nowSource: NVD / NIST