CVE-2019-25719

HIGHCVSS 8.8/10EPSS 0.13%

Last modified

CVE-2019-25719 is a high-severity vulnerability rated 8.8/10 on the CVSS scale. Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.. EPSS estimates a 0.13% chance of exploitation in the next 30 days.

Description

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.

Metrics

CVSS 3.1
8.6/10

CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H

CVSS 4.0
8.8/10

CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X

EPSS Probability
0.13%

3.1th percentile

Probability of exploitation in the next 30 days. Learn more

Weakness Enumeration

Affected Software

Source: CNA advisory (CVE.org). NVD analysis pending.

VendorProductVersions
DrägerInfinity Acute Care System<= VG4.1.1; <= VG4.0.3
DrägerStandalone Infinity M540 patient monitor<= VG4.1.1

References

Timeline

Published
Last Modified
Status
Awaiting Analysis

Frequently Asked Questions

What is CVE-2019-25719?
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.
How severe is CVE-2019-25719?
CVE-2019-25719 has a CVSS score of 8.8/10 (HIGH severity). The EPSS model estimates a 0.13% probability of exploitation in the next 30 days.
How do I fix CVE-2019-25719?
Check the vendor references and advisories linked above for patched versions and mitigation guidance. You can also run a Strix scan to test if your systems are affected.