CVE-2019-25723

MEDIUMCVSS 6.3/10EPSS 0.24%

Last modified

This CVE is reserved or awaiting analysis. Details will appear once published by NVD.

Description

Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.

Metrics

CVSS 3.1
4/10

CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:N/A:L

CVSS 4.0
6.3/10

CVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:N/VI:N/VA:L/SC:N/SI:N/SA:L/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X

EPSS Probability
0.24%

14.5th percentile

Probability of exploitation in the next 30 days. Learn more

Weakness Enumeration

References

Timeline

Published
Last Modified
Status
Awaiting Analysis

Are you affected by CVE-2019-25723?

Run a free Strix scan to check your systems for this vulnerability.

Scan your code now

Source: NVD / NIST