CVE-2019-25723

MEDIUMCVSS 6.3/10EPSS 0.24%

Last modified

CVE-2019-25723 is a medium-severity vulnerability rated 6.3/10 on the CVSS scale. Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.. EPSS estimates a 0.24% chance of exploitation in the next 30 days.

Description

Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.

Metrics

CVSS 3.1
4/10

CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:N/A:L

CVSS 4.0
6.3/10

CVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:N/VI:N/VA:L/SC:N/SI:N/SA:L/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X

EPSS Probability
0.24%

14.7th percentile

Probability of exploitation in the next 30 days. Learn more

Weakness Enumeration

Affected Software

Source: CNA advisory (CVE.org). NVD analysis pending.

VendorProductVersions
DrägerPerseus A5002.00; 2.01; 2.02

References

Timeline

Published
Last Modified
Status
Awaiting Analysis

Frequently Asked Questions

What is CVE-2019-25723?
Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.
How severe is CVE-2019-25723?
CVE-2019-25723 has a CVSS score of 6.3/10 (MEDIUM severity). The EPSS model estimates a 0.24% probability of exploitation in the next 30 days.
How do I fix CVE-2019-25723?
Check the vendor references and advisories linked above for patched versions and mitigation guidance. You can also run a Strix scan to test if your systems are affected.