CVE-2015-3953
Last modified
CVE-2015-3953 is a vulnerability of currently unknown severity. Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. EPSS estimates a 1.97% chance of exploitation in the next 30 days.
Description
Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Metrics
Weakness Enumeration
Affected Software
| Vendor | Product | Versions |
|---|---|---|
| Pifzer | Plum A\+ Infusion System Firmware | <= 13.4 |
| Pifzer | Plum A\+3 Infusion System Firmware | <= 13.6 |
| Pifzer | Symbiq Infusion System Firmware | <= 3.13 |
References
- https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01Mitigation, Third Party Advisory, US Government Resource
- https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01Mitigation, Third Party Advisory, US Government Resource
Timeline
- Published
- Last Modified
- Status
- Modified
Frequently Asked Questions
What is CVE-2015-3953?
How severe is CVE-2015-3953?
How do I fix CVE-2015-3953?
How Strix Helps
- How Strix found a critical auth bypass in etcdStrix autonomously discovered a critical authentication bypass in etcd, later designated CVE-2026-33413.
- Autonomous PentestingAI agents that find and validate exploitable vulnerabilities like this one across your applications.
- PR ReviewsPentest every pull request so vulnerable code is caught before it ships to production.
- AI Penetration TestingHow AI-driven penetration testing continuously covers your attack surface.
Related CVEs from 2015
- CVE-2015-3947SQL injection vulnerability in Advantech WebAccess before 8.…
- CVE-2015-3948Cross-site scripting (XSS) vulnerability in Advantech WebAcc…
- CVE-2015-3949Sinapsi eSolar Light with firmware before 2.0.3970_schsl_2.2…
- CVE-2015-3950Cross-site request forgery (CSRF) vulnerability in XZERES 44…
- CVE-2015-3951RLE Nova-Wind Turbine HMI devices store cleartext credential…
- CVE-2015-3952Wireless keys are stored in plain text on Hospira Plum A+ In…
- CVE-2015-3954Hospira Plum A+ Infusion System version 13.4 and prior, Plum…
- CVE-2015-3955Stack-based buffer overflow in Hospira LifeCare PCA Infusion…
- CVE-2015-3956Hospira Plum A+ Infusion System version 13.4 and prior, Plum…
- CVE-2015-3957Hospira LifeCare PCA Infusion System before 7.0 stores priva…
- CVE-2015-3958Hospira LifeCare PCA Infusion System 5.0 and earlier, and po…
- CVE-2015-3959The firmware in MNS before 4.5.6 on Belden GarrettCom Magnum…
Are you affected by CVE-2015-3953?
Run a free Strix scan to check your systems for this vulnerability.
Scan your code nowSource: NVD / NIST