CVE-2015-3956 is a vulnerability of currently unknown severity. Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. EPSS estimates a 0.95% chance of exploitation in the next 30 days.
Description
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Metrics
Weakness Enumeration
Affected Software
| Vendor | Product | Versions |
|---|---|---|
| Pifzer | Plum A\+ Infusion System Firmware | <= 13.4 |
| Pifzer | Plum A\+3 Infusion System Firmware | <= 13.6 |
| Pifzer | Symbiq Infusion System Firmware | <= 3.13 |
References
- https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01Mitigation, Third Party Advisory, US Government Resource
- https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01Mitigation, Third Party Advisory, US Government Resource
Timeline
- Published
- Last Modified
- Status
- Modified
Frequently Asked Questions
What is CVE-2015-3956?
How severe is CVE-2015-3956?
How do I fix CVE-2015-3956?
How Strix Helps
- Same Subject, Wrong User: A Cross-Issuer Account Takeover in n8nStrix found an identity-binding bug in n8n's token-exchange flow enabling account takeover.
- Autonomous PentestingAI agents that find and validate exploitable vulnerabilities like this one across your applications.
- PR ReviewsPentest every pull request so vulnerable code is caught before it ships to production.
- AI Penetration TestingHow AI-driven penetration testing continuously covers your attack surface.
Related CVEs from 2015
- CVE-2015-3950Cross-site request forgery (CSRF) vulnerability in XZERES 44…
- CVE-2015-3951RLE Nova-Wind Turbine HMI devices store cleartext credential…
- CVE-2015-3952Wireless keys are stored in plain text on Hospira Plum A+ In…
- CVE-2015-3953Hard-coded accounts may be used to access Hospira Plum A+ In…
- CVE-2015-3954Hospira Plum A+ Infusion System version 13.4 and prior, Plum…
- CVE-2015-3955Stack-based buffer overflow in Hospira LifeCare PCA Infusion…
- CVE-2015-3957Hospira LifeCare PCA Infusion System before 7.0 stores priva…
- CVE-2015-3958Hospira LifeCare PCA Infusion System 5.0 and earlier, and po…
- CVE-2015-3959The firmware in MNS before 4.5.6 on Belden GarrettCom Magnum…
- CVE-2015-3960The firmware in MNS before 4.5.6 on Belden GarrettCom Magnum…
- CVE-2015-3961The web-server component in MNS before 4.5.6 on Belden Garre…
- CVE-2015-3962Schneider Electric StruxureWare Building Expert MPM before 2…
Are you affected by CVE-2015-3956?
Run a free Strix scan to check your systems for this vulnerability.
Scan your code nowSource: NVD / NIST