CVE-2015-3954

UnknownEPSS 2.04%

Last modified

CVE-2015-3954 is a vulnerability of currently unknown severity. Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. EPSS estimates a 2.04% chance of exploitation in the next 30 days.

Description

Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

Metrics

EPSS Probability
2.04%

78.6th percentile

Probability of exploitation in the next 30 days. Learn more

Weakness Enumeration

Affected Software

VendorProductVersions
PifzerPlum A\+ Infusion System Firmware<= 13.4
PifzerPlum A\+3 Infusion System Firmware<= 13.6
PifzerSymbiq Infusion System Firmware<= 3.13

References

Timeline

Published
Last Modified
Status
Modified

Frequently Asked Questions

What is CVE-2015-3954?
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
How severe is CVE-2015-3954?
Severity scoring for CVE-2015-3954 is pending analysis. The EPSS model estimates a 2.04% probability of exploitation in the next 30 days.
How do I fix CVE-2015-3954?
Check the vendor references and advisories linked above for patched versions and mitigation guidance. You can also run a Strix scan to test if your systems are affected.

Are you affected by CVE-2015-3954?

Run a free Strix scan to check your systems for this vulnerability.

Scan your code now

Source: NVD / NIST